Precision You Can Trust
NABL-Accredited Calibration | Available on Request
The IWATSU CS-8200 Semiconductor Curve Tracer is a high-voltage semiconductor characterization mainframe from the CS-8000 Series, designed for precise testing of power semiconductor devices including MOSFETs, IGBTs, diodes, SiC and GaN devices. Its integrated high-voltage unit delivers up to 2,000 V peak at 20 mA, while the built-in medium-voltage unit supports up to 200 V / 1 A or 100 V / 2 A. With 250 fA minimum current measurement resolution, programmable gate control and a 12.1-inch touch display, the CS-8200 supports detailed static, leakage and high-voltage semiconductor characterization..
The IWATSU CS-8200 is designed for semiconductor R&D, power electronics laboratories, component manufacturers and engineering teams that require high-voltage current-voltage characterization with accurate low-current measurement.
The dedicated HV unit provides a maximum peak output of 2 kV at 20 mA, corresponding to a maximum peak output power of 40 W. It supports both DC and pulse operation, positive and negative polarity, and constant-voltage or constant-current drive measurements. The HV measured-voltage range extends from 50 V to 2 kV, while the normal measured-current range extends from 50 μA to 20 mA.
For leakage-current characterization, the HV measurement system also provides dedicated low-current ranges extending down to 5 nA, complementing the instrument's minimum measured-current resolution of 250 fA. This makes the CS-8200 suitable for breakdown and leakage evaluation as well as conventional high-voltage semiconductor characteristic measurements.
In addition to the HV source, the CS-8200 incorporates the CS-8000 Series medium-voltage unit. It supports peak operating combinations of 200 V / 1 A and 100 V / 2 A, with measured voltage ranges from 200 mV to 200 V and measured current ranges from 5 nA to 2 A. Supported waveforms include DC, pulse, sine, half-wave and full-wave operation.
The integrated GATE unit delivers up to 40 V / 1 A and supports DC, pulse and 50 Hz sine waveforms. Programmable gate sequencing, a -40 V to +40 V pre-signal gate range, and configurable gate hold times allow semiconductor devices to be evaluated under different bias conditions.
The CS-8200 also supports single sweep and double sweep measurements. Double-sweep operation is particularly useful for evaluating hysteresis behaviour in wide-bandgap devices such as SiC and GaN semiconductors.
A large 12.1-inch touch screen provides X-Y characteristic curves, Y-T time-domain waveform monitoring and measurement information. Y-T monitoring allows engineers to inspect applied waveforms, measurement points, pulse timing and abnormal behaviour such as oscillation during device characterization.
For applications requiring significantly higher current, the CS-8200 can be combined with optional CS-205, CS-210 or CS-220 high-current units, extending pulse-current capability to 500 A, 1,000 A or 2,000 A respectively at up to 50 V. IWATSU specifically shows a system configuration using the CS-8200 + CS-220 + CS-322 fixture for a 2 kV / 2,000 A characterization system.
PMK America's CS-8200 configuration requires the CS-021 HV standard connection cable set and CS-322 test fixture, together with the CC-8200 calibration certificate and ID-8200 inspection test data. Optional CS-2xx current-extension units, CS-520 patch panel and CS-5xx test adapters are not included with the basic mainframe.
| Product Type | Semiconductor Curve Tracer Mainframe |
| Series | CS-8000 Series |
| Model | CS-8200 |
| HV Unit | Integrated |
| Maximum HV Peak Voltage | 2,000 V |
| Maximum HV Current | 20 mA |
| Maximum HV Peak Power | 40 W |
| HV Measured Voltage Range | 50 V to 2 kV |
| HV Normal Measured Current Range | 50 μA to 20 mA |
| HV Leakage Current Ranges | Down to 5 nA |
| Minimum Measured Current Resolution | 250 fA |
| HV Waveforms | DC, Pulse |
| HV Polarity | Positive / Negative |
| HV Drive | Constant Voltage / Constant Current |
| HV Pulse Width | 20 ms to 1.6 s |
| MV Maximum Peak Output | 200 V / 1 A or 100 V / 2 A |
| MV Measured Voltage Range | 200 mV to 200 V |
| MV Measured Current Range | 5 nA to 2 A |
| MV Minimum Current Resolution | 250 fA |
| MV Waveforms | DC, Pulse, Sine, Half-Wave, Full-Wave |
| Gate Maximum Peak Output | 40 V / 1 A |
| Gate Measured Voltage Range | 1 V to 50 V |
| Gate Measured Current Range | 5 nA to 1 A |
| Gate Minimum Current Resolution | 250 fA |
| Gate Waveforms | DC, Pulse, Sine 50 Hz |
| Gate Pre-Signal Range | -40 V to +40 V |
| Gate Hold Time | 0 to 5 s |
| Measurement Cycle | 2 ms to 5 s |
| Measurement Modes | Single, Repeat, Stop |
| Sweep Modes | 1-Point, Single, Double |
| Sweep Steps | 1 to 1,000 |
| Display | 12.1-inch TFT LCD Touch Panel |
| Display Resolution | 1280 × 800 WXGA |
| Display Modes | X-Y, Y-T, Data List, Cursor |
| LAN Interface | 10/100/1000BASE-T |
| External Storage | USB Memory |
| Power Supply | AC 100 V to 240 V ±10%, 50/60 Hz |
| Maximum Power Consumption | 220 VA |
| Dimensions | 424 W × 221 H × 556 D mm |
| Weight | Approx. 21 kg, excluding accessories/options |
| Performance Temperature Range | +10°C to +35°C |
| Operating Temperature Range | 0°C to +40°C |
| Warm-Up Time | 30 minutes |
IWATSU specifies the CS-8200 as the 2 kV / 20 mA mainframe in the CS-8000 family. Its MV and gate sections are common to the series, while the HV section distinguishes it from the CS-8020.
| Model | Maximum Peak Current | Maximum Voltage |
| CS-205 | 500 A | 50 V |
| CS-210 | 1,000 A | 50 V |
| CS-220 | 2,000 A | 50 V |